Newheek manufacture high quality X Ray Grids.Welcome to purchase.
Email

HomeBlog ›What are the specifications of the grid?

What are the specifications of the grid?

Do you know what are the specifications of the grid? The main function of the X-ray grid is to filter out scattered rays. Scattered rays are due to the fact that when X-rays pass through the human body, a part of them has a special effect with human tissue, so that the original X-rays change the direction of propagation and form rays with different propagation directions called scattered rays. X-ray grids are mainly used for film frames, film beds, gastrointestinal beds, etc.
(1) Grid variable size
From standard 5×7” to 14×51”, special sizes can also be customized according to customer requirements, and the size of the X-ray grid is one inch larger than the corresponding film size
(2) Variable grid ratio of the grid
From 3:1 to 15:1, the grid ratio refers to the ratio of the height of the lead grid bar to the width of the gap
(3) The focal length range of the grid
From standard short focus (26”-32”) medium focus (34”-44”) long focus (40”-72”, 48”-72”60”-72”) to infinity (parallel lines)
(4) High grid density grid
From 60 or 70 threads/inch, to high density (85 threads/inch), super high density (103 threads/inch), invisible threads (105 threads/inch), super invisible threads (170 or 200 threads/inch)
If you have other questions about the grid, please call us. Whatsapp:+86 18953679166. Email: service@newheek.com .

X-ray grid

Author:X Ray Grids Maker


(+86) 18653679166
service@newheek.com