Basic knowledge of DR imaging equipment maintenance: no image after exposure
This fault can be dealt with in two ways. The first is to check whether the generator emits radiation. If not, check the generator side. For example, whether the hand switch is broken, whether the generator is faulty (you can open the generator software separately to check), whether there is a problem with the tube. If the generator can be exposed normally, you should check the plate side. For example, whether the connection of the tablet is normal (whether the network cable or optical fiber is connected correctly), or whether the synchronization signal (exposure enable signal) of the tablet is normal. Find the fault from these two aspects, you can find the fault.